Moves to a specified position and ejects a specified amount from the needle.
Parameter | Value |
Description |
---|---|---|
Eject |
Default | The default amount specified in the Set Up Standard Autosampler panel. |
Maximum | The maximum amount defined by the injector configuration. | |
<value> (µL) | A specified amount that overrides the default amount. | |
to |
Sample | The location specified in the sample information or sequence/worklist. |
Location | A specified location. | |
Seat | The seat capillary. | |
Speed |
Default | The default speed specified in the Set Up Standard Autosampler panel. |
Maximum | The maximum eject speed. | |
<value> (µL/min) | A value that overrides the default eject speed. | |
useOffset |
Default | The default offset position, specified in the Set Up Standard Autosampler panel. |
<value> (mm) | -2.5 to +35 mm below or above the normal offset position |
The parameters shown are dependent on the selected source.