Moves the autosampler needle as specified.
Parameter | Value |
Description |
---|---|---|
Up | Moves the needle to its upper position. | |
Down | Moves the needle to its lower position. | |
Into | Moves the needle into a specified Source with a specified Offset.The Source can be Sample, Location or Vial+. | |
Sample | The location specified in the sample information or sequence/worklist with a specified Offset. | |
Location | A specified location. | |
Offset (mm) |
Def. | The default draw position, specified in the Setup Injector dialog box. |
Value | -2.5 to +35 mm below or above the normal draw position The offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample. |
The parameters shown are dependent on the selected source.