Select the appropriate volume for the syringe from the drop-down list.
Select the volume of the front seat capillary from the drop-down list. Agilent provides seat capillaries with the following volumes:
2.3 µL, 5 µL, 6 µL, 20 µL, and 24 µL
Select the volume of the back seat capillary from the drop-down list. Agilent provides seat capillaries with the following volumes:
2.3 µL, 5 µL, 6 µL, 20 µL, and 24 µL
Enter a loop volume in the field, either by using the up- and down-arrows or by typing the volume in the field. You can enter a volume between 0.00 and 50000.00 µL, in increments of 0.01 µL.
Enter a loop volume in the field, either by using the up- and down-arrows or by typing the volume in the field. You can enter a volume between 0.00 and 50000.00 µL, in increments of 0.01 µL.
This check box is marked to indicate that a BCD/external contacts board has been detected during autoconfiguration. For manual configuration, mark the External contacts board installed check box to indicate that a BCD/external contacts board is installed. Specify the use of external contacts in the External Contacts.
When the External contacts board installed check box is cleared, the External Contacts section of the method setup is not present. |
use BCD port for and BCD port output format are used to configure the BCD output, which is available only if you have installed the BCD/external contacts board.
Select Location to transfer the location information in BCD format. For example, you can connect an integrator which supports BCD to transfer the location information to your integrator reports.
Select Binary output if you want to control third party devices using binary commands sent by the Output command of the Autosampler Pretreatment.
This check box is marked to indicate that a thermostat has been detected during autoconfiguration. For manual configuration, mark the Thermostat installed check box to indicate that a thermostat is installed.
This check box is for use in offline mode only. In online mode, use the Modify command from the Dashboard context menu to chenge the thermostat control.
Mark the check box to specify that the thermostat is controlled by the parameters in the Thermostat and Enable Analysis sections of the Advanced method parameters. In this case, the thermostat parameters in the Control dialog box are deactivated.
Clear the check box to specify that the thermostat is controlled by the Thermostat Temperature Control and Enable Analysis sections of the Control dialog box. In this case the parameters in the Method dialog box are deactivated. This is the default condition.
Displays the Define and edit Wellplates dialog box, which contains a list of standard preconfigured wellplates plus any custom wellplates that have been added.