Wash Needle (High Performance Autosampler Control)

The Wash Needle dialog box allows you to wash the needle without modifying your current method. The needle wash specified in this dialog box is carried out immediately.

Mode

Select the wash location by clicking the down-arrow and selecting the needle wash mode from the list. You can select either Flush Port (default location) or Wash Vial.

Time

This field is enabled only if you have selected Flush Port. It enables you to specify a flush time in seconds between 0.0 and 100.0 in steps of 0.1. The peristaltic pump continues purging while the needle is lifted to ensure that the wash solvent is flushed completely to waste.

Location

This field is enabled only if you have selected Wash Vial. The Location field allows you to specify a wash location.

See Specifying a Location.

Repeat

This field is available only if you have selected Wash Vial. You can specify up to 5 repeat washes; the default is 3.

Draw Position

This field is available only if you have selected Wash Vial. It enables you to specify the needle height in mm while drawing solvent; you can specify values between -10.0 and 50.0 in steps of 0.1