This section is available only when a rinse valve is configured and enabled. It is not available when a G4227A Flexible Cube is configured. |
Mark the Enable Rinse check box to enable the rinse parameters. When the check box is cleared, the rinse parameters are disabled.
The Rinse Draw Speed determines the speed at which the plunger draws rinse solvent from the vial.
Type | Syringe (μL) | Min (μL/min) | Max (μL/min) |
---|---|---|---|
High Performance Autosampler | 20 |
4.0 | 1000 |
High Performance Autosampler | 40 | 4.0 |
1000 |
High Performance Autosampler |
100 | 10 | 2500 |
High Performance Autosampler | 900 |
90 | 2500 |
μ Wellplate sampler | 8 | 0.7 |
100 |
μ Wellplate sampler |
40 | 0.7 | 250 |
The Rinse Eject Speed determines the speed at which the plunger ejects sample from the metering device.
Limits
See Draw Speed above
Enter the rinse volume of organic solvent as a multiple of the maximum draw volume (syringe + loop capillary + seat capillary).
Enter the rinse volume of water as a multiple of the maximum draw volume (syringe + loop capillary + seat capillary).