Moves the autosampler needle as specified.
Parameter | Value |
Description |
---|---|---|
Up | Moves the needle to its upmost position. | |
Down | Moves the needle to its lowest position. | |
Into | Moves the needle into a specified Source with a specified Offset. The Source can be Sample, Location, Seat, Upmost, Sample+ or Vial+. The Offset of the autosampler needle is calculated from the BOTTOM of the well, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the well bottom with the needle. Use a positive offset to raise the needle higher in the well to clear precipitates on the bottom of the well, or to select one phase of a biphasic sample. The unit is in millimeters, with 1/10 mm resolution. The range of offsets available is from -10 to 50 mm. | |
Sample | The position specified in the sample information or sequence/worklist. | |
Location | From a specified location. The location can be a well or a vial. | |
Seat | The seat capillary. | |
Upmost | The upper position of the needle.This command is equivalent to using the Up parameter. | |
This function is for vial trays only. | ||
The parameters shown are dependent on the selected source.