The Drawer tab of the SFC Multisampler configuration displays the current configuration read from a connected instrument. It is not intended for use in changing this configuration. To change the configuration of drawers, capillaries or the reference vial, open the context menu of the SFC Multisampler dashboard and choose the desired function from the Modify submenu.
The Drawer tab can be used for an offline configuration. However such settings will be changed as soon as an instrument is connected. For offline configuration, for each position (starting at the bottom), click the down-arrow and select a drawer type from the drop-down list. You can configure drawers to a total of eight height units (HU), for example, eight 1H drawers or four 2H drawers. If more than eight HU are requested, an error is shown.
Start the configuration from the bottom of the list, and add the drawers as they would be inserted in the instrument. |
The offline drawer configuration does not change an existing online configuration. |