Auto-clean Settings and Start (SFC Multisampler)

Use this dialog box to specify the parameters for the auto-clean function. The auto-clean parameters are in two sections:

Injection Valve

This section is not available in SFC operating mode.

Switch Injection Valve

Mark the check box to specify that the injection valve is switched during the auto-clean process so that both paths are cleaned.

Standard Wash

Wash Needle

Mark the check box to specify that the needle is washed. Specify the wash time (in seconds) in the adjacent Duration field.

Flush Flow Path

Mark the check box to specify that flow path of the sampler is flushed with solvent delivered by the main pump of the system.

In LC operating mode, specify the flush time (in seconds) in the adjacent Duration field. In SFC operating mode the duration is set by the firmware.

Click Start to start the auto-clean procedure.

Stop a running auto-clean procedure by selecting Stop Auto-clean from the context menu of the status panel in the Dashboard.