Sample Flush-Out Factor |
This parameter is available in LC operation mode only. The Sample Flush-Out Factor ensures that the sample is thoroughly flushed out of the syringe and past the capillary seat and valve. The factor is part of the flush-out volume formula, which is calculated by the SFC Multisampler firmware. The volume is calculated as: factor x (injection volume + seat capillary + valve volume). The Sample Flush-Out Factor is preset to 5.0 at the factory. The preset Sample Flush-Out Factor is correct for most methods. However, for unusually viscous samples, you should increase the Sample Flush-Out Factor to obtain the desired degree of flushing to prevent sample carry-over. |
Injection Valve to Bypass for Delay Volume Reduction |
This parameter is available in LC operation mode only. The parameter is used to switch the flow from the injector from the main flow path to bypass after injection has taken place. This reduces the delay volume for low volume techniques. You can specify the point during the analyses when the valve switches to bypass. This is done by setting the Sample Flush-out Factor. |
Enable Overlapped Injection | Overlapped injection provides faster throughput of samples by allowing the preparation of the next injection while the current injection is being completed. The net effect is more samples performed during a given period of time, although the actual individual injection times remain constant. When overlapped injection mode is employed, the next sample is taken up into the SFC Multisampler as the first sample is being injected. Choose the When sample is flushed out option (available in LC operation mode only) to specify that the sample can be prepared directly after the current sample has been flushed out. Choose the After Period of Time option to specify that the next sample is prepared after a user-defined period of time. This is the only available option in SFC operation mode. The parameters for overlapped injection are specified as part of the method. There are no parameters for overlapped injection to be specified in the sequence/worklist table. However, there are important restrictions to using Overlapped Injection in sequences/worklists:
|