Enables the Multisampler to draw from a specified Source.
Parameter |
Value |
Description | |||
---|---|---|---|---|---|
Draw | Default | The default amount specified in the current method. See Set Up Multisampler. | |||
Maximum | The maximum amount defined by the multisampler configuration. | ||||
<Value> | A value that overrides the injection volume specified in the current method. | ||||
From |
Sample | The default position as specified in the sample information or sequence/worklist. | |||
Location | A specified location on one of the sample containers that are loaded in the hotel. The location can be a well or a vial. | ||||
Current Container | A user-specified location that refers to the currently used sample container, allowing you to omit the multisampler drawer to be used from the specification. In this case, it is sufficient to specify the row and column of the location to be used. | ||||
Sample+ | A user-specified increment to the previously specified vial location (as specified in the sample information or sequence/worklist) by N positions; sample is drawn from the new position.
| ||||
Location+ | A user-specified increment to the previously specified wellplate location (as specified in the sample information or sequence/worklist) by N rows and M columns; sample is drawn from the new position.
| ||||
Plate+ | A user-specified increment to the previously specified plate position (as specified in the sample information or sequence/worklist) by N positions; sample is drawn from the new plate. | ||||
Air | Draws in air through the needle. The action is carried out at the current needle position. No further movement of the needle is carried out | ||||
Seat | From the seat capillary. (Not supported for Multi-wash or Dual-needle operation) | ||||
Speed | Default | The default draw speed as specified in the current method. See Set Up Multisampler. | |||
Maximum | The maximum draw speed. | ||||
<Value> | A value that overrides the draw speed specified in the current method. | ||||
Use Offset |
Default | The default draw position as specified in the current method. See Set Up Multisampler. | |||
<Value> | A value that overrides the Needle Height Position Offset specified in the current method. |