In the Flush Solvent Bottle dialog box, specify the amount of liquid in the flush solvent bottle, and specify the action to be taken when the level gets too low.
The parameters are in two sections:
In the Filling section, specify the capacity of the flush solvent bottle, and the current volume in the bottle. The flush solvent bottle has two fields:
Enter the capacity of the flush solvent bottle (in liters) in the Total Volume field. The maximum value is 100.00.
Enter the amount of the solvent currently in the flush solvent bottle (in liters) in the Actual Volume field. The maximum value of the Actual Volume is the Total Volume of the bottle.
The Actual Volume field is updated as the flush solvent is used. When you fill the flush solvent bottle, enter the new actual volume of the flush solvent bottle in the appropriate Actual Volume field.
No direct measurement of the levels in the flush solvent bottle is done; the amount remaining is calculated by the amount of flush solvent used, and the Actual Volume set. |
In the Actions section, specify the action to be taken if the flush solvent level falls too low.
Mark the Prevent analysis if level falls below check box, and enter a threshold in the field to ensure that the pump generates a Not Ready condition if the calculated flush solvent bottle filling level falls below the set threshold. The current run is completed, but the sequence/worklist will not continue with the next analysis.
Mark the Generate error if running out of flush solvent check box to ensure that the sampler generates an error if the calculated flush solvent bottle filling level is zero. The error is recorded in the logbook and the analysis is stopped immediately.