Options (Online Sample Manager/Hybrid Multisampler)

The contents of this tab are set automatically during autoconfiguration. For offline usage, you can set the parameters manually.

Metering

Shows the currently installed metering device. For offline usage, click the down arrow to select a different metering device.

Left Needle installed

This function is not applicable to the Online Sample Manager or Hybrid Multisampler; the dual-needle option is not supported for this device.

Right Needle

Loop

Shows the currently installed loop capillary. For offline usage, click the down arrow to select a different loop capillary.

Seat

Shows the currently installed seat capillary. For offline usage, click the down arrow to select a different seat capillary.

Pressure Sensor installed

This check box is marked to show that a Pressure Sensor is installed.

Thermostat installed

This check box is marked to show that a sample cooling device or thermostat is installed.

  • Mode

    Select Constant temperature mode to set the temperature using the Thermostat section of the Online Sample Manager or Hybrid Multisampler Control parameters. Use this mode to store samples at a constant temperature across multiple runs. This is the default mode, which is recommended in most cases.

    Select Variable temperature mode to set the temperature using the Enable Analysis section of the Advanced Method Setup parameters. In this mode, the temperature can be varied from run to run.

    Use Variable temperature mode with care. All samples in the storage compartment are exposed to the set temperature for any current method. Unintended change of the sample temperature may cause degradation. It may take some time to reach the set temperature.

  • Thermostat

    Click the down-arrow and select the installed device (Cooler or Thermostat) from the drop-down list.

    The Online Sample Manager or Hybrid Multisampler Cooler does not offer a heating function. A temperature higher than the previous setting, or higher than ambient, may only be reached after a long time, if ever.

Multi-wash SSV v.2

(G3167A only)

This check box is marked to show that the following Hydraulic box Multiwash Assembly is installed:

  • G3167-68601

    This hydraulic box does not support Solvent S2 for outer wash.

If the check box is not marked, the following Hydraulic box Multiwash Assembly is installed.

  • G3167-68600

    This hydraulic box supports Solvent S2 for outer wash.

Reference Vial Rack

Click the down-arrow and select the reference vial rack that is installed in your Online Sample Manager or Hybrid Multisampler from the drop-down list.

Bypass capillary

This function is not applicable to the Online Sample Manager or Hybrid Multisampler; the bypass capillary is not available for this device.

External Sampling Valve Settings (G3167A/B only)

Shows the currently installed External Sampling Valve head. For offline usage, click Configure to configure a different External Sampling Valve head.

Supported External Sampling Valve heads:

  • G3167A works only with 3-pos/6-port valve 800 bar (5067-6680).

  • G3167B works only with 3-pos/6-port bio valve 1300 bar (5320-0003).

The External Sampling Valve is not supported for G7167C.

Define Sample Containers

Displays the Define and edit Wellplates dialog box, which contains a list of standard preconfigured wellplates plus any custom wellplates that have been added.