Missing Vessel (Online Sample Manager/Hybrid Multisampler Control)

Ignore Missing Vessel

The injector ignores a missing vessel and continues with a 6-second dummy run. The message Missing vessel <x> is logged, and the system continues with the next injection.

Abort Current Sequence

The injector generates an error and stops the analysis. The message Missing vessel <x>, sequence aborted is logged, and the sample or sequence/worklist is aborted.

Abort Current Run

The injector generates an error and stops the run. The message Missing vessel <x>, run skipped is logged, and the run is skipped.

The setting applies to standard injections only. If a missing vessel is encountered in an injector program, the current Missing Vessel setting is ignored, and the whole sequence is aborted.

Limitation