Feed Injection (Online Sample
Manager)
This section is available only when Feed
injection mode has been selected. The default values depend on the selection in
Method Preset.
Feed Speed |
The Feed Speed is the
speed of delivery of the sample into the solvent flow.
Adaptive Select
this option to set the Feed Speed to a
proportion of the pump flow. Typically, a high value (default 80%) is used for
isocratic analyses; and a low value (default 5%) is used for gradient
analyses. Fixed Select this
option to set the Feed Speed to a fixed value; enter the value in the
field. |
Flush
Out |
Specifies the volume of solvent to use to flush out the injection
path. Mode Mark the
automatic check box to use the volume of
solvent calculated based on the injection volume and the volumes of the seat
and loop capillaries.
Volume
When the automatic
check box is cleared, specify the flush-out volume in the
field. |