Classical Flow-through Mode (Injection Path Cleaning)

Inner Wash Mode

Adds the following steps to the table. These steps are carried out before the sample is drawn:

Off

No Inner Wash is performed.

Standard

  • Inner wash

    Select the solvent to use and specify the volume.

  • Reconditioning

    Solvent S2 is used to flush the injection path.

Extended

  • Inner wash

    Select the solvent to use for the first wash and specify the volume.

  • Inner wash

    Select the solvent to use for the second wash and specify the volume.

  • Seat wash

    Select the solvent to use for the first seat wash and specify the volume.

  • Seat wash

    Select the solvent to use for the second seat wash and specify the volume.

  • Reconditioning

    Solvent S2 is used to flush the injection path.

Outer Wash Mode

Adds the following steps to the table. These steps are carried out after the sample is drawn, before the sample is injected:

Off

No Outer Wash is performed.

Standard

  • Outer wash

    Select the solvent to use and specify the duration of the wash.

Extended

  • Outer wash

    Select the solvent to use for the first wash and specify the duration of the wash.

  • Outer wash

    Select the solvent to use for the second wash and specify the duration of the wash.