Modify Temperature Mode (Autosampler Control)

Click the down-arrow and select the temperature mode from the drop-down list.

Constant temperature mode (control setting)

Select Constant temperature mode to set the temperature using the Thermostat section of the Control parameters. Use this mode to store samples at a constant temperature across multiple runs. This is the default mode, which is recommended in most cases.

Variable temperature mode (method parameter)

Select Variable temperature mode to set the temperature using the Thermostat section of the Advanced Method Setup parameters. In this mode, the temperature can be varied from run to run.

Use Variable temperature mode with care. All samples in the storage compartment are exposed to the set temperature for any current method. Unintended change of the sample temperature may cause degradation. It may take some time to reach the set temperature.

Note that the Cooler does not offer a heating function. A temperature higher than the previous setting, or higher than ambient, may only be reached after a long time, if ever.