Enables the autosampler to draw from a specified source.
Parameter | Value |
Description | |||
---|---|---|---|---|---|
Draw |
Default | The default amount specified in the Set Up Vialsampler panel. | |||
Maximum | The maximum amount defined by the injector configuration. | ||||
<value> (µL) | An amount that overrides the default injection volume. | ||||
from |
Sample | From the position specified in the sample information or sequence/worklist. | |||
Location | From a specified location. | ||||
Seat | From the seat capillary. | ||||
Air | From air, above the seat capillary. | ||||
Sample+ | A user-specified increment to the previously specified location (as specified in the sample information or sequence/worklist) by N positions; sample is drawn from the new position.
| ||||
Location+ | A user-specified increment to the previously specified location (as specified in the sample information or sequence/worklist) by N rows and M columns; sample is drawn from the new position.
| ||||
Speed |
Default | The default speed, specified in the Set Up Vialsampler panel. | |||
Maximum | The maximum draw speed | ||||
<value> (µL/min) | A value that overrides the default draw speed. | ||||
useOffset |
Default | The default draw position, specified in the Set Up Vialsampler panel. | |||
<value> (mm) | -2.5 to +35 mm below or above the normal draw position The offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample. |