Column Compartment Dashboard Panel

Module Graphic

The items in the column compartment graphic have the following meaning and function, depending on the configured options:

Shows the status of the column thermostat, off

or on.

Shows the current position of the valve. The graphic shows the actual installed valve.

Not all valves are available for all modules.

The column graphic tooltip shows the column tag details (if available).

The current temperature(s) of the column thermostat(s) are displayed below the column graphic.

Actuals

The following column thermostat actuals are displayed:

Temp. Left

The current actual temperature (left) and setpoint temperature (right) of the left column heat exchanger.

Temp. Right

The current actual temperature (left) and setpoint temperature (right) of the right column heat exchanger.

Valve Position

    The current valve position:

  • 6-port valve

    Ports 1 & 2 or ports 1 & 6.

  • 10-port valve

    Ports 1 & 2 or ports 1 & 10.

Context Menu

The context menu contains the following commands, depending on the configured options:

Control

Displays the column compartment's Control dialog box.

Method

Displays the column compartment's Method Setup dialog box.

Error Method

  • Set Error Method

    Sets the error method to the method that is currently available in the hardware. The error methods are loaded automatically on all modules if an error occurs.

  • Clear Error Method

    Removes the currently loaded error method.

Identify Device

Causes the LED on the front of the module to blink for a few seconds.

Switch Valve

Displays the column compartment's Switch Valve Position dialog box.

Switch On

Switches the column compartment thermostat on.

Switch Off

Switches the column compartment thermostat off.

Modify

(G1316C only)

    Choose what to modify from the submenu:

  • Valve type

    Displays the column compartment's Modify Valve Type dialog box, which allows you to register the replacement of a valve with an old RFID tag by a valve with a new tag.