FLD Dashboard Panel

Module Graphic

The items in the FLD graphic have the following meaning and function:

Indicates that an External Contacts board is installed.

The the status of the xenon lamp, off (left) or on (right)

When balancing is in progress, the graphic is yellow (not ready).

Actuals

The FLD actuals shows a table of the current signals in four columns:

Signal

The signal definition.

ExWL

The excitation wavelength (in nm).

EmWL

The emission wavelength (in nm).

LU

The signal value (in LU).

In addition, the PMT Gain below the signals table shows the current photomultiplier gain setting.

Context Menu

The context menu contains the following commands:

Control

Displays the detector's Control dialog box.

Method

Displays the detector's Method Setup dialog box.

Error Method

  • Set Error Method

    Sets the error method to the method that is currently available in the hardware. The error methods are loaded automatically on all modules if an error occurs.

  • Clear Error Method

    Removes the currently loaded error method.

Identify Device

Causes the LED on the front of the module to blink for a few seconds.

Balance

Balances the detector.

Switch on

Ignites the xenon lamp.

Switch off

Switches off the xenon lamp.

Take 3D Scan

(G1321A/B and G7121B only)

Acquires a 3D scan, which plots the measured intensity against the excitation and emission wavelengths.