The items in the RID graphic have the following meaning and function:
|
The RID is in standard mode, neither purging nor recycling. |
The RID is purging; the purge valve is on. | |
The RID is recycling; the recycle valve is on. | |
|
When balancing is in progress, the graphic is yellow (not ready). |
Displays the state of the optical unit heater, off (left) or on (right). | |
Indicates that an External Contacts board is installed (G1362A only). |
The following RID actuals are displayed:
Optical Unit Temperature |
The current temperature of the optical unit (in °C). |
Diode Balance |
The diode balance ratio between sample and reference. |
Recycling | The status of the recycling valve (either off or the length of time it has been on). |
Purge |
The status of the purge valve (either off or the length of time it has been on). |
RI Signal | The signal output (in nRIU). |
Polarity | The current signal polarity (pos or neg). |
Wait after Purge |
The time remaining after purging before the RID is ready. |
The context menu contains the following commands:
|
Displays the detector's Control dialog box. |
|
Displays the detector's Method Setup dialog box. |
|
|
| Causes the LED on the front of the module to blink for a few seconds. |
|
Balances the detector. |
|
Switches the optical unit heater on. |
|
Switches the optical unit heater off. |
|
Opens the purge valve. |
|
Closes the purge valve. |
|
Switches on recycling of the solvent. |
| Switches off recycling of the solvent. |