Lens1 DC (LC/MS)

The Lens1 DC voltage controls the voltage applied to Lens1. Lens 1 DC affects ions of all masses in a similar manner. An incorrect Lens1 DC setting will result in no, or greatly reduced, ion transmission.

Minimum Maximum Default Typical
-5.0 5.0 2.0 +2.0 (positive ions) -2.0 (negative ions)

During manual tuning, the Lens1 DC voltage can be ramped to find an optimum value. Ramp defaults and limits are:

  Default (volts) Limits (volts)
Start (From) -5 -5 to (Stop - Step)
Stop (to) 5 (Start + Step) to 5
Step Size (by) 0.1 0.1 to (Stop - Start)

See Also

Tune > Optics Section

Tune the MS manually (LC/MS)