MS DC (LC/MS)

The MS DC parameter controls a dc offset voltage applied to the quadrupole mass filter.

This voltage affects peak widths and ion transmission. It has its greatest effect on the transmission of higher mass ions.

Polarity Minimum Maximum Default Typical
Positive 0 20 3.0 0.0 to 5.0
Negative -20 0 -3.0 -5.0 to 0.0

During manual tuning, the MS DC voltage can be ramped to find an optimum value, but ramping the MS DC voltage is not a reliable way to find an optimum value. Ramping the MS DC is used primarily as a diagnostic tool.

Ramp defaults and limits are:

Positive Polarity

  Default (volts) Limits (volts)
Start (From) 0 0 to (Stop - Step)
Stop (to) 13 (Start + Step) to 13
Step Size (by) 0.1 0.1 to (Stop - Start)

Negative Polarity

  Default (volts) Limits (volts)
Start (From) -13 -13 to (Stop - Step)
Stop (to) 0 (Start + Step) to 0
Step Size (by) 0.1 0.1 to (Stop - Start)

See Also

Tune > Optics Section

Tune the MS manually (LC/MS)