2.19 Selector stream valve cannot switch properly.

Probable Causes

  • Stream valve is bad

  • Bad valve PCA

Suggested Actions

  • Replace stream valve.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages