2.20 AC line zero cross circuit has failed. Vent and power cycle is needed to clear condition.

Probable Causes

  • Bad AC board

  • Bad AC power line

  • Bad AC line filter

Suggested Actions

  • Power cycle.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument