Online Signals Window

During a run you may want to monitor some of the parameters in the run. Some of these parameters can change during the run.

You can monitor the following SQ parameters:

  • *TIC: Scan - A TIC scan over the mass range in that segment in the method.

  • Rough Vac - the pressure in the region between the capillary and the skimmer.

  • High Vac - the pressure in the region of the quadrupole and the detector.

  • Turbo1 Speed

  • Pump1 Current

  • MS Heater - the temperature of the Quad Mass Filter. The default temperature is 100 °C. You cannot change this temperature.

  • *Gas Temperature - See Gas Temperature.

  • *Gas Flow - See Gas Flow.

  • *Nebulizer - See Nebulizer.

  • * Capillary Voltage - See Capillary Voltage.

  • *Chamber Current

  • *Capillary Current - See Capillary Current.

  • *Corona Voltage -

  • *Vaporizer - See Vaporizer.

  • APPI Lamp - See Turn APPI lamp on or off .

*The parameter can change during the run.