Preparations
You need the following privileges. Privileges are configured in the Control Panel. |
|
In Data Analysis, load the result set with your acquired oligo data.
Create a new method with method configuration MS, and save it.
Link the new method to the result set.
Set the method parameters as specified below.
Reprocess and save your results.
Method section | Settings |
|---|---|
|
General > Signals , Blank Subtraction tab |
Blank subtraction applied on: Select All injections Use specific blank: Select the third blank in your current result set. Ensure that the third blank injection baseline closely resembles the baseline of the 25 µL sample You have to set this parameter individually for each new result set. |
|
Extraction > Chromatogram , MS tab | Single m/z expansion chromatograms: This is typically set to either 0.3, 0.7 asymmetrical, or ±0.5 symmetrical. Determine this parameter empirically. |
|
Extraction > Spectrum MS tab |
|
|
Compounds > Calibration , General Table tab |
Set Number of levels to 4. |
|
Compounds > Identification , Compound Table tab |
Add a compound for your full-length product (FLP). Use exactly the same name and spelling as in the ion list.
|
|
Compounds > Calibration , Compound Table tab |
|