This topic provides a procedure to change from a splitless injection using a split/splitless inlet to a solvent vent mode injection using a Multimode Inlet (MMI). It applies mainly to large volume injections (LVI) using a MMI, but the concepts can apply to general MMI use. This topic does not consider all items that can impact an analysis, for example the liner, solvent, analyte boiling points, or polarity. This topic also assumes knowledge of your data system—when to save a method, how to start a run, how to set up a single run, etc.
The main advantage of the MMI solvent vent mode is that you can inject slowly into the inlet, allowing large amounts of solvent to be evaporated from the liner while concentrating the analytes inside the liner. This requires an injector with variable speed injections, a "large" syringe, and knowledge of the sample and the solvent. When developing a solvent vent method, the goal is to determine the injection rate, inlet temperature, inlet flow, and inlet pressure needed to evaporate selectively the solvent but not the analytes. The development technique is first to calibrate the system for analyte recovery and then scale up the conditions to achieve the desired method improvement. The most significant parameters are:
Inlet temperature. Hold the inlet temperature at or below the solvent boiling point until after all the sample has been injected. This is important so that you do not vent the more volatile analytes along with the solvent. Solvent vent methods work best when the boiling point of the solvent is at least 100 °C lower than the first eluting analyte. This allows an unpacked liner to be used which gives the most inert inlet possible. Smaller boiling point differences can be handled but typically require some type of retentive packing in the liner (e.g. Tenax).
Injection speed. If using a MMI, use the Solvent Elimination Calculator. This calculator requires the solvent type, injection volume, and boiling point of the first eluting analyte to determine the correct injection speed. Normally, the injection speed is set so that the rate of injected liquid solvent just matches the rate of eliminated vaporized solvent. This establishes a concentrating zone inside the liner to retain the analytes of interest.
Vent time. Make sure the vent time setpoint is greater than the time the needle spends in the inlet. If the vent time is too short, you will inject more solvent than necessary. This can lead to distorted peak shapes and too much solvent reaching the detector. As you increase the injection volume, the vent time will also increase to eliminate more solvent.
Vent flow. The vent flow rate is directly proportional to the solvent vent rate. The vent flow is set to be high enough to remove the solvent in a reasonable time without removing the analytes. For high vent flow rates, the inlet backpressure will increase. This may require the method to control the vent pressure above ambient in order to maintain method precision.
Vent pressure. The vent pressure affects how easily the solvent vaporizes and how much enters the column during solvent evaporation. The lower the vent pressure, the more quickly the solvent is removed and the less solvent that enters the column. Vent flow can impact vent pressure in the following way. As the vent flow increases, the actual vent pressure minimum will also increase. For best method precision, the vent pressure should be set slightly above ambient pressure so that the instrument reliably controls it.
Rijks and coworkers developed the theory behind solvent elimination in these inlets. The following simplified equation is used to calculate the solvent elimination rate from the parameters described above.
Solvent elimination rate = C x Pvap/Tinlet x Po/Pi x F
Where:
C = solvent related constant
Pvap = vapor pressure of solvent at inlet temperature
Tinlet = inlet temperature
Po = outlet pressure
Pi = inlet vent pressure
F = inlet vent flow
If using an MMI with an MSD, another tip for method development is to scan for solvent ions. Detecting the solvent ions can be useful in determining how much residual solvent is being transferred to the column.
In most cases, you should see five times the peak area for the analytes. If the response of all analytes is too low:
If the response of early eluters is too low: