This check box is marked to indicate that a BCD/external contacts board has been detected during autoconfiguration. For manual configuration, mark the External contacts board installed check box to indicate that a BCD/external contacts board is installed. Specify the use of external contacts in the External Contacts.
| When the External contacts board installed check box is cleared, the External Contacts section of the method setup is not present. | 
This check box is marked to indicate that a thermostat has been detected during autoconfiguration. For manual configuration, mark the Thermostat installed check box to indicate that a thermostat is installed. Specify the thermostat parameters in the Fraction Collector Control dialog box.
Mark this check box to indicate that a UIB is connected, and an MSD is connected to it, to enable mass-based fraction collection.
| The MSD is not detected automatically; you must mark the check box manually. |