The items in the pump graphic have the following meaning and function:
| 
 | (G1361A only) Indicates that an External Contacts board is installed. | 
| 
 
 
 | When the solvent level falls below the specified volume, the yellow warning is shown; when the bottle is empty, the red error is shown. Clicking on the solvent bottle displays the Bottle Fillings dialog box. The tooltip for the bottle shows the solvent name. | 
| 
 | (G7161B only) Indicates the presence of a solvent selection valve. Use the context menu to switch the valve. | 
| 
 | The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). | 
| 
 | (G7161B with second pressure sensor only) The pressure setpoints. The red lines show the current maximum pressure limits; the gray areas show the current pressures (also shown as text above and below the graphic). | 
The current solvent flow rate (in mL/min) is displayed above the pressure display.
The following pump actuals are displayed:
| Flow | The current solvent flow rate (in mL/min). | 
| Pressure | The current pump pressure (in bar or psi). | 
| Pressure A, B | (G7161B with second pressure sensor only) The current pressure readings of each of the two pressure sensors. | 
| Ripple | (G1361A and G7161B) The current ripple (in %). | 
| Ripple A, B | (G7161B with second pressure sensor only) The current ripple readings (in %) of each of the two pressure sensors. | 
| Pressure Limit | The current maximum pressure limit. The reason for any limitation to the maximum pressure limit is displayed (G7161A/B only) | 
| Flow Limit | The current maximum flow limit. The reason for any limitation to the maximum flow limit is displayed (G7161A/B only) | 
| Composition A, B | (G7161A/B only) The percentage contribution of each of the solvent channels to the current solvent composition. | 
| Purge Status | (G1361A only) The current status of the purge, on or off. | 
| Purge Time | (G1361A only) The current purge time (in minutes). | 
The context menu contains the following commands:
| 
 | Displays the pump's Control dialog box. | 
| 
 | Displays the pump's Method Setup dialog box. | 
| 
 | 
 | 
| 
 | Causes the LED on the front of the module to blink for a few seconds. | 
| 
 | Toggles the status of the pump, on or off. | 
| 
 | (G7161B only) Switches the solvent selection valve to the alternative position. | 
| 
 | Displays the Bottle Fillings dialog box. | 
| 
 | (G1361A only) Toggles the status of the purge, on or off. | 
| 
 | (G7161A/B only) Allows you to ensure that all air bubbles in the seal wash lines are removed (for example, after the seal wash solvent bottle has been refilled). | 
| 
 | 
 |