The items in the pump graphic have the following meaning and function:
| 
 | Indicates that an External Contacts board is installed. | 
| 
 
 
 | When the solvent level falls below the specified volume, the yellow warning is shown; when the bottle is empty, the red error is shown. Clicking on the solvent bottle displays the Bottle Fillings dialog box. The tooltip for the bottle shows the solvent name. | 
| 
 
 | (G4204A, G7104A/C and G7131A/C) Indicates that the ISET option is installed but with no active method (gray) or installed and active (black). | 
| 
 | (G4204A, G7104A/C and G7131A/C) Indicates that Blend Assist is activated in the method. | 
| 
 | The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). | 
The current solvent flow rate (in mL/min) is displayed above the pressure display.
The following pump actuals are displayed:
| Flow | The current solvent flow rate (in mL/min). | 
| Pressure | The current pump pressure (in bar or psi). | 
| Ripple | (Not G4204A or G7104A/C) The current ripple (in %). | 
| Tuning | (G4204A and G7104A/C) Indicates the tuning efforts of 1290 Infinity pumps. For pumps operating as expected, the signal should stay in a range of -1 to +1 within the full scale of -2 to +2. | 
| Pressure Limit | The current maximum pressure limit. | 
| Composition A, B, C, D | The percentage contribution of each of the solvent channels to the current solvent composition. | 
| Mixer | The installed mixer type. | 
| Solvent <n> | The solvent configured for each channel. | 
| Purge Status | The current status of the pump purge. (G4204A only) | 
| Prepare Pump | Displays the current pump preparation operation and the remaining duration for Purge or Conditioning, or the run time for Prime. (G7104A/C only) | 
| Valve Position | The current valve position. (G4204A and G7104A/C) | 
The context menu contains the following commands:
| 
 | Displays the pump's Control dialog box. | 
| 
 | Displays the pump's Method Setup dialog box. | 
| 
 | 
 | 
| 
 | Causes the LED on the front of the module to blink for a few seconds. | 
| 
 | Toggles the status of the pump, on or off. | 
| 
 | Displays the Bottle Fillings dialog box. | 
| 
 | (G4204A only) Toggles the status of the pump purge, on or off. | 
| 
 | (G4204A only) Turns on the pump priming function. Proming is turned off automatically. | 
| 
 | (G4204A only) Toggles the status of system conditioning, on or off. | 
| 
 | (G7104A/C and G7131A/C) Displays the Prepare Pump dialog box. | 
| 
 | (G4204A, G7104A/C and G7131A/C) Toggles the status of filter flushing, on or off. | 
| 
 | (G7104A/C and G7131A/C) Allows you to ensure that all air bubbles in the seal wash lines are removed (for example, after the seal wash solvent bottle has been refilled). |