1.13 Octopole RF driver cannot maintain the requested RF peak voltage.

Probable Causes

  • Octopole RF of Octopole PCA not dipped

  • Shortage in manifold

  • Bad main board

Suggested Actions

  • If you have removed or cleaned the octopole assembly, verify that all cables are connected correctly.

  • Check rods and cables.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages