1.17 MS RF driver cannot maintain the requested mass setting.

Probable Causes

  • Quad not dipped

  • Overheated Quad driver

  • Broken Quad driver

  • Shortage in manifold

  • Broken RRPA

  • Broken 28V PS

Suggested Actions

  • If you have cleaned or changed the quadrupole, then verify that all cables are connected correctly.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument