1.24 HED power supply senses a fault condition.

Probable Causes

  • Bad HED power supply

  • Bad cable

  • Shortage to ground in manifold

Suggested Actions

  • Check the vacuum.

  • Turn the HED off from the Detector tab in the Tune > Manual Tune Section (LC/MS).

  • Check for leak.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument