1.3 Octopole shroud driver cannot maintain the requested voltage.

Probable Causes

  • Main board power supply broken

  • Ribbon cable disconnected and shorted to ground

  • Octopole PCA broken

  • Skimmer 1 and Skimmer 2 lens shorten together (6120C)

Suggested Actions

  • If the octopole assembly has recently been removed, then verify that it is installed correctly.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument