1.31 Leak detected in leak tray

Probable Causes

  • Solvent leak in leak tray

  • Bad leak sensor

  • Bad valve PCA

Suggested Actions

  • Check for leak and wipe off solvent if there is liquid.

  • Power cycle to see if the error goes away.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument