1.5 Lens2 DC driver cannot maintain the requested voltage.

Probable Causes

  • Mainboard problem

  • Problem with main board lens cable

  • Lens1 and Lens2 shorten together

Suggested Actions

  • If the octopole assembly has recently been removed, then verify that it is installed correctly.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument