2.10 Cable to the octopole board is not connected. Turn off power to the electronics before connecting cable.

Probable Cause

G1946-60080 cable not connected or bad contact

Suggested Actions

  • Turn off power and reconnect cable. When you connect the cable to the valve board, do not force it, or the pins will be damaged.

  • If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages

Tune the instrument