System timed out after sensing temperature fault
Bad drying gas heater/sensor assembly
Bad AC board
Power cycle the instrument and monitor the drying gas heater read back and set point. Set the Gas Temp to 300 °C. If the heater is not working, the actual value for the Gas Temp (in the Method > Source Section (LC/MS)) will not climb. It will remain between ambient and 150 °C. If the sensor is shorted, it may read much less than 10 °C.
If this error continues to occur, copy this error on the screen and contact Agilent for more help.