View Scan spectra in Tune section

Scans are visible in the Spectrum Window when you are acquiring data.

Enter the Dwell time (ms).
  1. Click Method.

    Ribbon_MethodLayout.jpg
  2. In the Acquisition Method window in the navigation pane, click Instrument Setup > SQ.

  3. If needed, click Spectrum.

    Ribbon_SelectSpectrum.jpg
  4. In the navigation pane, select Tune > Manual Tune.

  5. In the navigation pane, select Tune > Manual Tune > Acquisition.

  6. If needed, click Take. In this image, Take is already selected.

    Ribbon_TakeControl.jpg
  7. Click TB_TuneControl.jpg to lock the instrument for tuning. You need to unlock the instrument after you finish tuning.

  8. Select the Polarity. Select Positive or Negative.

  9. Select Scan for the Acquisition type.

  10. Enter the Dwell time (ms).

  11. Select the Stepsize.

  12. Enter the Samples and Averages for Scan.

  13. Select or clear Centroid.

  14. Enter the Threshold.

  15. Click TB_AcquireWithTuneParameters.jpg.

    In the Spectrum window, an MS scan is shown. You can continue to do other actions in Tune such as ramping parameters and changing values while you are acquiring data.

  16. When you have finished acquiring data, click TB_StopTuneAcquisition.jpg.

  17. Click TB_UnlockTune.jpg to unlock the instrument. You need to unlock the instrument after you finish tuning.

See Also

View Profile scans

Tune the MS manually

Tune the MS using Fast Scan

Tune > Acquisition Section