Octopole RF (LC/MS)

The Octopole RF parameter controls the peak-to-peak (ac) voltage applied between the poles of the Octopole.

Minimum

Maximum

Default

0 V

800 V

600 V

During manual tuning, the Octopole RF voltage can be ramped to find an optimum value. Ramp defaults and limits are:

Minimum

Default (volts)

Limits (volts)

Start (From)

0.0

0 to (Stop - Step)

Stop (to)

800

(Start + Step) to 800

Step Size (by)

25.0

5.0 to (Stop - Start)