Calculations for relative retention times in 2 dimensions

Calculations are always done using the same dimension for RRT and expected RT.

Update of RRT for expected 2D compounds in the 2D-LC processing method

RRT ¹D = Expected ¹D RTcompound / Expected ¹D RTreference

and

RRT ²D = Expected ²D RTcompound / Expected ²D RTreference

RRT of 2D peaks listed in the Injection Results window

For identified compounds:

RRT ¹D = Actual ¹D RTcompound / Actual ¹D RTreference

and

RRT ²D = Actual ²D RTcompound / Actual ²D RTreference

For other 2D peaks (including not identified compounds), if one or more 2D RRT reference compounds exist:

RRT ¹D = Actual ¹D RTpeak / Actual ¹D RT1st reference

and

RRT ²D = Actual ²D RTpeak / Actual ²D RT1st reference

See Also

Calculations for relative retention times in 1 dimension

×