The items in the pump graphic have the following meaning and function:
| Indicates that an External Contacts board is installed. |
| When the solvent level falls below the specified volume, the yellow warning is shown; when the bottle is empty, the red error is shown. The tooltip for the bottle shows the solvent name. If Level Sensing is configured, clicking on the solvent bottle displays the Level Sensing Assignment dialog box. Otherwise the Bottle Fillings dialog box is displayed. |
| (G4220A/B, G7120A and G7132A) Indicates that the ISET option is installed but with no active method (gray) or installed and active (black). |
| Indicates the presence of a solvent selection valve. Use the context menu to switch the valve. |
| The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). |
The current solvent flow rate (in mL/min) is displayed above the pressure display.
The following pump actuals are displayed:
Flow | The current solvent flow rate (in mL/min). |
Pressure | The current pump pressure (in bar or psi). |
Ripple | (Not G4220A/B, G7120A or G7132A) The current ripple (in %). |
Tuning <Channel> | (G4220A/B, G7120A and G7132A) Indicates the tuning efforts of 1290 Infinity pumps. For pumps operating as expected, the signal should stay in a range of -1 to +1 within the full scale of -2 to +2. |
Pressure Limit | The current maximum pressure limit. |
Composition A, B | The percentage contribution of each of the solvent channels to the current solvent composition. When a solvent selection valve is fitted, the channels are shown in the graphic. |
Purge Status | The current status of the pump purge. (G4220A/B only) |
Prepare Pump | Displays the current pump preparation operation and the remaining duration for Purge or Conditioning, or the run time for Prime. (G7120A and G7132A only) |
Mixer | The installed mixer type. (G4220A/B, G7120A and G7132A) |
The context menu contains the following commands:
Control | Displays the pump's Control dialog box. | |
Method | Displays the pump's Method Setup dialog box. | |
| Error Method |
|
| Identify Device | Causes the LED on the front of the module to blink for a few seconds. |
Switch On/Off | Toggles the status of the pump, on or off. | |
Switch Solvent Selection Valve A/B | Switches the solvent selection valve to the alternative position. These menu items are present only when a solvent selection valve is configured. | |
Bottle Fillings | Displays the Bottle Fillings dialog box. | |
Level Sensing Assignment | Displays the Level Sensing Assignment dialog box. The menu item is present only when Level Sensing is configured. | |
Zero | Displays the Zero dialog box. The menu item is present only when Level Sensing is configured. | |
| Purge On/Off | Allows you to control the purging of the system. (G4220A/B only) |
| Prime On/Off | Allows you to prime the pump heads if they are full of air. (G4220A/B only) |
| Conditioning On/Off | Allows you to switch pump conditioning on and off. See Conditioning for more information. (G4220A/B only) |
| Prepare Pump | Displays the Prepare Pump dialog box. (G7120A and G7132A only) |
| Seal Wash Prime | Allows you to ensure that all air bubbles in the seal wash lines are removed (for example, after the seal wash solvent bottle has been refilled). (G7120A and G7132A only) |