The items in the high performance autosampler graphic have the following meaning and function:
Displayed during an injector program, either internal or from the injector program/pretreatment table. The highlight moves during the injection. At the end of the run, the icon is removed. | ||
Displays the status of the sample thermostat. This graphic is available only when a sample thermostat is installed.
| ||
Indicates that an External Contacts board is installed. | ||
Indicates that a G4227A Flexible Cube is installed. | ||
The syringe graphic shows a representation of the flowpath
| ||
Denotes an injection from a wellplate. The current sample location is shown to the right of the graphic. The tooltip displays the current tray configuration. |
The following high performance autosampler actuals are displayed:
Injection Volume | The volume of the last injection (in µL). |
Sample Location | The location of the last sample. |
Temperature | The sample thermostat temperature (in ° C). This actual is present only if a sample thermostat is installed. |
The context menu contains the following commands:
Control | Displays the sampler's Control dialog box. | |
Method | Displays the sampler's Method Setup dialog box. | |
| Injector Program | Displays the sampler's Injector Program. |
| Error Method |
|
| Identify Device | Causes the LED on the front of the module to blink for a few seconds. |
Home Arm | Moves the needle arm to its home position to allow you to remove the tray. | |
Reset Sampler | Sends a reset signal to your autosampler. During the reset, the injector is in a Not Ready state (yellow). | |
Wash Needle | Displays the sampler's Wash Needle dialog box, which allows you to wash the needle without modifying your current method. | |
Maintenance Position | Moves the arm to the front of the autosampler and the needle to its uppermost position. | |
| Switch Valve to Bypass | Switches the valve from the main flow path to bypass. |
| Switch Valve to Mainpass | Switches the valve from bypass to the main flow path. |
| Edit Wellplate Types | Displays the Wellplates dialog box. |
Assign Wellplates | Displays the Tray and Plate Configuration dialog box. | |
| Flexible Cube | Choose an action from the submenu:
|
| Modify | Temperature Mode Displays the sampler's Modify Temperature Mode dialog box, which allows you to switch between Constant temperature mode, with control from the Control parameters, and Variable temperature mode, with control from the Method Setup parameters. Use Constant temperature mode to store samples at a constant temperature across multiple runs. This is the default mode, which is recommended in most cases. In Variable temperature mode, the temperature can be varied from run to run. |