Select the setting to use if a vessel is missing:
Ignore Missing Vessel |
The injector ignores a missing vessel and continues with a 6-second dummy run. The message Missing vessel <x> is logged, and the system continues with the next injection. |
Abort Current Sequence |
The injector generates an error and stops the analysis. The message Missing vessel <x>, sequence aborted is logged, and the sample or sequence/worklist is aborted. |
Abort Current Run |
The injector generates an error and stops the run. The message Missing vessel <x>, run skipped is logged, and the run is skipped. |
NOTE
The setting applies to standard injections only. If a missing vessel is encountered in an injector program, the current Missing Vessel setting is ignored, and the whole sequence is aborted.
Not all versions of Chromatography Data Systems (CDS) support Abort Current Run. Selecting this option will trigger an Abort Current Sequence.
base-id: 3269461387