The items in the pump graphic have the following meaning and function:
| Indicates that an External Contacts board is installed. |
| When the solvent level falls below the specified volume, the yellow warning is shown; when the bottle is empty, the red error is shown. Clicking on the solvent bottle displays the Bottle Fillings dialog box. The tooltip for the bottle shows the solvent name. |
| Indicates the presence of a solvent selection valve. Use the context menu to switch the valve. |
| The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). |
The following pump actuals are displayed:
Flow | The current solvent flow rate (in mL/min). |
Pressure | The current pump pressure (in bar or psi). |
Ripple | The current ripple (in %). |
Pressure Limit | The current maximum pressure limit. |
Composition A, B | The percentage contribution of each of the solvent channels to the current solvent composition. When a solvent selection valve is fitted, the channels are shown in the graphic. |
Primary Flow | The currently specified primary flow. |
EMPV Duty Cycle | The percentage of the time that the EMPV has been on. |
Purge Status | The current status of the pump purge. |
Purge Channel | The current purge channel. |
Purge Time | The current purge time (in minutes) |
The context menu contains the following commands:
Control | Displays the pump's Control dialog box. | |
Method | Displays the pump's Method Setup dialog box. | |
| Error Method |
|
| Identify Device | Causes the LED on the front of the module to blink for a few seconds. |
Switch On/Off | Toggles the status of the pump, on or off. | |
Switch Solvent Selection Valve A/B | Switches the solvent selection valve to the alternative position. | |
Bottle Fillings | Displays the Bottle Fillings dialog box. | |
| Purge on/off | Switches on and off system purging as specified in the Pump Configuration. |
| Start/Stop Fast Composition Change | Starts or stops the fast composition change as specified in the Pump Configuration. |
base-id: 3443999883