The items in the pump valve cluster graphic have the following meaning and function:
| Indicates that an External Contacts board is installed. |
| When the solvent level falls below the specified volume, the yellow warning is shown; when the bottle is empty, the red error is shown. Clicking on the solvent bottle displays the Bottle Fillings dialog box. The tooltip for the bottle shows the solvent name. |
| (G4220A/B, G4204A, G7120A, G7132A and G7104A/C) Indicates that the ISET option is installed but with no active method (gray) or installed and active (black). |
| The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). |
The current solvent flow rate (in mL/min) is displayed above the pressure display.
The following pump actuals are displayed:
Flow | The current solvent flow rate (in mL/min). |
Pressure | The current pump pressure (in bar or psi). |
Ripple | (Not G4220A/B, G7120A, G7132A, G4204A or G7104A/C) The current ripple (in %). |
Tuning | (G4220A/B, G7120A, G7132A, G4204A and G7104A/C) Indicates the tuning efforts of 1290 Infinity pumps. For pumps operating as expected, the signal should stay in a range of -1 to +1 within the full scale of -2 to +2. |
Pressure Limit | The current maximum pressure limit, and the module that is setting the limit. |
Composition A, B | The percentage contributions of solvent channels A and B to the current solvent composition. |
Composition C, D | The percentage contributions of solvent channels C and D to the current solvent composition. |
Valve Position | The current valve position. |
Purge Status | The current status of the pump purge. |
Prepare Pump | Displays the current pump preparation operation and the remaining duration for Purge or Conditioning, or the run time for Prime. (G7120A, G7132A and G7104A/C) |
The context menu contains the following commands, depending on the configured options:
Control | Displays the pump's Control dialog box. | |
Method | Displays the pump's Method Setup dialog box. | |
| Error Method |
|
| Identify Device | Displays a submenu that allows you to identify either each individual device or all devices in the cluster: Selecting an individual device from the submenu causes the LED on the front of the selected device to blink for a few seconds. Selecting All from the submenu causes the LEDs of all devices in the cluster to blink. |
Switch On/Off | Switches the pump on when it is off, or off when it is on. | |
Bottle Fillings | Displays the Bottle Fillings dialog box. | |
| Purge On/Off | Allows you to control the purging of the system. (G4220A/B and G4204A only) |
| Prime On/Off | Allows you to prime the pump heads if they are full of air. (G4220A/B and G4204A only) |
| Conditioning On/Off | Allows you to switch pump conditioning on and off (G4220A/B and G4204A only) See Conditioning for more information. |
| Prepare Pump | Displays the Prepare Pump dialog box (G7120A, G7132A and G7104A/C only) |
| Seal Wash Prime | Allows you to ensure that all air bubbles in the seal wash lines are removed (for example, after the seal wash solvent bottle has been refilled). (G7120A, G7132A and G7104A/C only) |