You can set limits on the rate of change of the solvent flow to protect your analytical column. Limits for up ramps and down ramps are set separately.
Limits:
|
G1361A: |
|
|
Up |
0.100 – 800.000 mL/min² in steps of 0.001 mL/min² |
|
Down |
0.100 – 10000.000 mL/min² in steps of 0.001 mL/min² |
|
G7161A/B: |
|
|
Up |
0.10 – 1000.000 mL/min² in steps of 0.1 mL/min² |
|
Down |
0.10 – 1000.000 mL/min² in steps of 0.1 mL/min² |
base-id: 3443578891