To Create a New HS Method Based on a Specific Application
To create a new method based on a known application, use the built-in
wizard. The wizard will apply changes to the currently-loaded method.
- If desired, save the current method. The wizard changes the current
method, so any unsaved changes will be lost.
- Configure the HS for the new method. See To
Configure the Headspace Sampler.
- Open the HS method editor.
- Select Method Development from the
navigation tree.
- Click Create method based on a specific application.
The wizard dialog appears.
- Select the sample's matrix type, Liquid
or Solid.
- Use the drop-down menu to select the sample vial size for the analysis.
- Enter the amount of sample to be processed.
- For liquid matrix samples, enter the volume of sample (in mL)
in the sample vial from into the Sample
Volume field.
- For solid matrix samples, enter the total sample mass (in g)
into the Sample Mass field.
- For the Solvent, use the drop-down
list to select the solvent used to prepare the sample.
- If using no solvent, for example, for solid samples, select
No Solvent.
- If using a solvent not shown in the list, select Other
Solvent. A Boiling Point field
appears. Enter the solvent's boiling point, in °C.
- A selected solvent's Boiling Point
displays for reference.
- In the Compound(s) of Interest section,
enter the highest boiling point (in °C) found in the compounds of
interest present in the sample.
- Click Preview Changes and review the
new and changed parameters.
- If acceptable, click Accept Changes.
The software will apply recommended settings to the current method.
These settings represent reasonable starting values during method
development. To return to the wizard, click Reject
Changes. See Confirm Changes
Dialog.
- Review the method.
- Change any parameter as needed.
- Save the revised method with a new name (if desired).
See also Creating a New Method Based on
a Specific Application.