The following procedure describes how to quantify all not identified peaks by using either Manual Factor or reference peaks.
Preparations
|
Under Methods in the navigation pane, select the processing method you want to edit.
Create a timed group with the following properties (see Create a timed group):
The time range covers the entire chromatogram.
The Include identified peaks check box is cleared.
The Quantify each peak individually check box is selected.
In the Compounds > Calibration node, select the required quantification mode.
To use a Manual Factor:
Under Mode, choose Manual Factor.
Provide a value in the Manual Factor column.
To use reference peaks:
Under Mode, choose Reference.
Under Curve reference, select the reference compound.
Provide a correction factor in the Ref. correction column.