Follow this procedure to set the correct number of calibration levels. This is especially important for one-point calibrations to avoid warnings on reprocessing.
Prerequisites
To be able to carry out the procedure as described, you need the privilege Processing Method > Edit calibration parameters. Privileges are configured in the Control Panel.
If the project enforces method approval, the method status must be Generic. For details, see Method approval.
If the Processing Method window is hidden, click Processing Method in the Home ribbon tab to show the window.
In the Processing Method window, select the Compounds > Calibration node.
Select the General tab.
Under Number of levels, enter the required number of calibration levels.
Select the Compound table tab.
The number of available calibration levels corresponds to the number you have entered in the previous step.